CPAP Set-Up

NIPPV 3 machines are used throughout the trust to deliver NIV and CPAP – and should be commenced in ED if transfer to ward/ICU is adding significant delay

  • NIV/CPAP  is an Aerosol Generating Proceedure (AGP)
    • Staff must wear full APG PPE
    • In Side room with a door
    • Door Marked with APG sign – HERE
  • CPAP/EPAP levels of 8-15cmH2O

This video demonstrates how to set up CPAP on the NIPPV 3

 

NIV (Non Invasive Ventilation)

NIV should be considered for use in patients with a  persisting Acute Hypercapnic Respiratory Failures after a maximum of one hour of standard medical therapy.

However, ICU should be contacted early if the patient has one of the following:

  • Asthma – Intubation the option of choice in Life threatening
  • Pneumonia – NIV should only be considered as a bridge to intubation
  • No pre-exisiting respiratory issue – NIV not likely helpful
  • pH <7.25 (low threshold for ICU input)
  • pCO2 >6.5kPa (low threshold for ICU input)
  • Type 1 Respiratory Failure (low threshold for ICU input)